Tyne Engineering’s new Model 7045 Room Tritium In-Air Monitor is a wall or rackmount version of our popular Model 7043 Portable Tritium Monitor. The 7045 is intended for room monitoring for both tritium and gamma. Tyne的新型7045室氚空气监测器是我们流行的型号7043便携式氚监测器的墙或机架式版本。7045的目的是为氚和伽玛的房间监测。
Specifications
| Measuring chamber effective volume | 1.5 L, (2 X 0.75L) | 
| Pump | Positive displacement gas pump: 12 VDC, flow rate can be up to 11L/min, continuous pressure 8 psig. | 
| Tritium measuring range | 0.2µCi/m3 (7,400Bq/ m3) to 1Ci/m3 | 
| 4 chambers | cruciform configuration: two for measuring and two for compensation | 
| Radiation measured | measuring total tritium, HTO, gamma and alpha | 
| Sensitivity | 0.2µCi/m3 (0.01 DAC) | 
| Accuracy | 0.4µCi/m3 from 0 to 100µCi/m3; ±10% from 100µCi/m to 1Ci/m3; | 
| Zero stability | ±0.4µCi/m3; | 
| Gamma measurement | Gamma is measured simultaneously with tritium. Measuring range is from 0.1mR/h to 10 R/hr; | 
| Radon compensation and tritium discrimination | Instrument can provide radon compensation and discriminate in a field of HTO and elemental H3 using the silica gel drier. | 
| Gamma background compensation | less than ±10% or 5mCi/m3 reading change whichever is greater response to 20mR/Hr; | 
| Ion trap | built into each ion chamber; | 
| Flow rate | 0 to 11 L/min | 
| Response time | 2 to 5s with 0-15 m air hose, within 20s with hose from 100 to 150m at 90% of signal | 
| Stability | : ±3% per 24 hours, ±10% per 30 days; | 
| Analog Output | 0-3V linear. | 
| Alarms | Tritium, gamma, air flow, detector failure, micro processor alarms in field as well as on optional remote display system through analog output (4 to 20mA) | 
| Alarm Response | < 30S with 2-150m hose; | 
| Purge decontamination | Chambers can be heated by a cartridge heater controlled by the unit to assist in purging, | 
| Condensation Prevention | With heater cartridge temperature can be raised above ambient to prevent condensation in instrument | 
| Communication port | RS232 (short distance) and RS485-bus (long distance); | 
| Data Logging frequency | The tritium value can be logged every minute and can be time and date stamped. (Gamma is not logged) | 
| Displays | One (1) TFT LCD large touch screen display (size 2.8’’), displays the values of tritium, gamma in both numerical and graphical format; one LED screen ¾’’ letters, 7 letter screen. | 
| Power supply | 220VAC (±10%); Power consumption < 1A | 
| Drier | silica gel cylinder is provided with the unit; | 
| Temperature operation range | 0 to 50ºC degree; | 
| Humidity operation range | 0-90% Non-condensable | 
| Size | Approx 9″High x 12″Deep x 19″ Wide; | 
| Weight | Approx.: 55lb; | 
| Mounting | Unit is designed for table-top, 19” rack, or wall-mounting (bracket optional) | 