7043 Portable Tritium-in-Air-Monitor 便携式测氚仪

Compensation for Gamma background is crucial in tritium measurement, since a gamma field of 1mR/Hr will generate 500 times the ionization generated in 1µCi/m³ of tritium. Tyne designed the four, 250 cc ion –chambers in a cruciform with each chamber connected to an ultra sensitive electrometer amplifier and filter. All signals pass through the ADC converter, and the micro-processor calculates and displays the tritium value. To cover high ranges of tritium a separate small ion chamber is used. Measurements such as flow from the solid-state flowmeter and chamber temperature are also monitored by the micro-processor. 对γ背景的补偿是氚测量的关键,因为1 mR/HR的伽玛场产生的电离量是1 mI/m氚电离的500倍。TYNE设计了四,250毫升离子室在一个十字形,每个室连接到一个超灵敏的静电计放大器和滤波器。所有信号通过ADC转换器,微处理器计算并显示氚值.为了覆盖高范围的氚,使用了一个单独的小离子室。测量结果,如来自固态流量计的流量和室内温度,也由微处理器进行监测。

Specifications

Sensitivity 1µCi/m³ ( 0.05 DAC)
Accuracy ±3µCi/m³ from 1 to 100µCi/m³, ±10% from 100µCi/m³ to 20,000 µCi/m³
Range 1 µCi/m³ to 199,999 µCi/m³
Detector 4 matched chambers – Two for measuring, two for compensation
Measuring chamber effective volume 500 cc (2 measuring chambers of 250cc each)
Flow rate 1.0 L/m
Zero stability ±1µCi/m³ immediately after unit is powered on
Background cancellation Less than 10% of the reading change in 20 mR/hr field
Radon compensation Included in the software
Ion trap Built into each ion chamber
Dust filter Built into the unit.
Noble gas cancellation Ports for air inlet/outlet of measuring chambers and compensation chambers are accessible on surface plate so silica gel drier can be placed between the chambers to compensate for presence of noble gas.
Tritium Discrimination Instrument can discriminate in a field of HTO and elemental H3 using the silica gel drier.
Purge/Decontamination All 4 Chambers can be heated up by a cartridge heater controlled by unit
Display Graphic touchscreen LCD display. Unit displays the values(tritium and Gamma) both in numerical format and graphical trend.
Communication Can be configured to RS232(Short distance) or RS485(Long distance).
Analog output 0-5 V Linear or logarithmic voltage output.
Power supply 2.5 V- 5V external power supply, or 3 C-cell battery. Total power consumption is 150mA.
Gamma measurement Gamma is measured directly by the unit. Gamma measuring range is from 0.1mR/h to 10R/h
Alarm Tritium alarms, gamma alarms, low air flow alarms will all be shown by red LED light and by an audible device.
Data logging The tritium and gamma values are logged every minute with current time and date stamp
Size and weight 8″(L)x5″(H)x5″(W), 7 lbs.
Drier Note that the silica gel drier is not part of the standard equipment, but can be supplied as an extra.
Response Time 30 seconds at 90% of signal