7043 Portable Tritium-in-Air-Monitor 便携式测氚仪

Compensation for Gamma background is crucial in tritium measurement, since a gamma field of 1mR/Hr will generate 500 times the ionization generated in 1µCi/m³ of tritium. Tyne designed the four, 250 cc ion –chambers in a cruciform with each chamber connected to an ultra sensitive electrometer amplifier and filter. All signals pass through the ADC converter, and the micro-processor calculates and displays the tritium value. To cover high ranges of tritium a separate small ion chamber is used. Measurements such as flow from the solid-state flowmeter and chamber temperature are also monitored by the micro-processor. 对γ背景的补偿是氚测量的关键,因为1 mR/HR的伽玛场产生的电离量是1 mI/m氚电离的500倍。TYNE设计了四,250毫升离子室在一个十字形,每个室连接到一个超灵敏的静电计放大器和滤波器。所有信号通过ADC转换器,微处理器计算并显示氚值.为了覆盖高范围的氚,使用了一个单独的小离子室。测量结果,如来自固态流量计的流量和室内温度,也由微处理器进行监测。

Specifications

Sensitivity1µCi/m³ ( 0.05 DAC)
Accuracy±3µCi/m³ from 1 to 100µCi/m³, ±10% from 100µCi/m³ to 20,000 µCi/m³
Range1 µCi/m³ to 199,999 µCi/m³
Detector4 matched chambers – Two for measuring, two for compensation
Measuring chamber effective volume500 cc (2 measuring chambers of 250cc each)
Flow rate1.0 L/m
Zero stability±1µCi/m³ immediately after unit is powered on
Background cancellationLess than 10% of the reading change in 20 mR/hr field
Radon compensationIncluded in the software
Ion trapBuilt into each ion chamber
Dust filterBuilt into the unit.
Noble gas cancellationPorts for air inlet/outlet of measuring chambers and compensation chambers are accessible on surface plate so silica gel drier can be placed between the chambers to compensate for presence of noble gas.
Tritium DiscriminationInstrument can discriminate in a field of HTO and elemental H3 using the silica gel drier.
Purge/DecontaminationAll 4 Chambers can be heated up by a cartridge heater controlled by unit
DisplayGraphic touchscreen LCD display. Unit displays the values(tritium and Gamma) both in numerical format and graphical trend.
CommunicationCan be configured to RS232(Short distance) or RS485(Long distance).
Analog output0-5 V Linear or logarithmic voltage output.
Power supply2.5 V- 5V external power supply, or 3 C-cell battery. Total power consumption is 150mA.
Gamma measurementGamma is measured directly by the unit. Gamma measuring range is from 0.1mR/h to 10R/h
AlarmTritium alarms, gamma alarms, low air flow alarms will all be shown by red LED light and by an audible device.
Data loggingThe tritium and gamma values are logged every minute with current time and date stamp
Size and weight8″(L)x5″(H)x5″(W), 7 lbs.
DrierNote that the silica gel drier is not part of the standard equipment, but can be supplied as an extra.
Response Time30 seconds at 90% of signal