7000 Tritium Controller 氚控制器

Tyne’s Tritium Controller is the main data processing centre for many of our tritium monitoring configurations. It is suitable for in-line monitoring of process systems, glove box monitoring, stack monitoring, room air monitoring, and other applications such as with Tyne’s tritium-in-breath analyzer. Tyne的氚控制器是我们许多氚监测配置的主要数据处理中心。适用于过程系统的在线监测、手套箱监控、栈监控、室内空气监测等应用, 如与Tyne的氚呼吸分析仪。

It can be configured for a variety of tritium monitoring requirements, including inputs from one or two ion chambers of different sizes and sensitivities (see Tyne’s 1000cc ion chamber and 10cc ion chamber), multiple alarm configurations, and multiple range configurations.它可以配置为各种氚监测要求, 包括来自一个或两个不同大小和灵敏度的离子室的输入 (见Tyne的1000cc 离子室和10cc 离子室), 多个报警配置, 和多个范围配置。

特征

大型平板触摸屏
最多可控制两个探测器
每通道双报警, 本地和远程
一个触摸自动零点
4-20 mA、0-10V 或 RS232 输出
平滑的汽车范围横跨9年

Specifications

Note: Specifications depend on configuration and ion chamber selection.

Detection Sensitivity with 1000 cc Detector 1 µCi/m3
Detection Sensitivity with 10 cc Detector 0.1 mCi/m³
Instrument Response Time less than 10 s
Detection Range with 1000 cc Detector 0 to 10 Ci/m3
Detection Range with 10 cc Detector 0 to pure tritium
Flow Rates with 1000 cc Detector 1000 cc/s recommended, 2000 cc/sec max
Flow Rates with 10 cc Detector 1000 cc/s
Operating Pressure Vacuum to 150 psig
Operating Temperature (ion chambers) 15 °C to 200 °C
Operating Temperature (controller) 0°C to 50°C
Operating Humidity 0% – 95% Relative Humidity
Leak Tightness less than 1×10-9 atm-cc/s He
Temperature Offset Under 1 µCi/°C
Gamma Offset 0.1 Ci/m3 per 1 R/h gamma fieldLinear response from 10-3 R/h to 102 R/h
Gamma Compensation Subtract with 2nd detector head
Background Compensation “One touch” Auto Zero
Tritium Recovery Chamber bakeable up to 350°C
Tritium Wetted Parts 316L Stainless Steel, High density Ceramic
Fittings VCR8
Controller Placement Up to 100m from detector head
Analogue Output 0 – 10 V
Current Output 4 – 20 mA
Digital Output RS232
Range Indication 3 digital output lines
Process Alarm Two alarm set points per channel
Process Control on Alarm Alarm Relay, 30 VAC/DC, dry contact relay
Compliance DOE Tritium Monitor Standard Rev 4 June 1999, ISO 9001:2000
Power Requirements 1A, 120 VAC, 50/60 Hz1/2A, 220 VAC 50/60 Hz
Dimensions of Controller Standard 19″ rack mount x 5″ Height
Dimensions of 1000 cc Detector 4.5″ diameter x 15″ length
Dimensions of 10 cc Detector 3″ diameter x 15″ length
Power 120 VAC 1A or 220 VAC 1/2A 50/60Hz