t2x2 Sport/EVO 支架 光学支架 精密温控样品池支架 t2双样品池支架

Dual temperature control for the Evolution 300 and 600 Spectrophotometers

t2x2 Sport/EVO

Dual temperature control for the Evolution 300 and 600 Spectrophotometers                                                         Evolution 300和600分光光度计的双温度控制

 

产品概述

The t2x2 Sport/EVO is an inexpensive, compact pair of cuvette holders providing independent Peltier temperature control for both the sample and reference cuvettes. Use it for temperature ranges of about -15 °C to 110 °C. It is ideal for most biological studies to observe absorbance spectra at precisely maintained temperatures. If necessary, run the sample and reference cuvettes at different temperatures. The t2x2 Sport/EVO consists of two separate cuvette holders, one for the sample and one for the reference, that mount in the spectrophotometer with utilities conveniently routed under the front panel of the instrument. Operate the TC 1/Dual Temperature Controller using the menu button or control via optional program T-App.

t2x2 sportyevo 是一种价格低廉、结构紧凑的一对试剂盒支架, 为样品和参考试剂盒提供独立的 peltier 温度控制。在-15°c 至110°c 的温度范围内使用。它是大多数生物研究的理想选择, 在精确保持的温度下观察吸收光谱。如有必要, 请在不同温度下运行样品和参考试剂盒。t2x2 sportyevo 由两个独立的 cuvette 支架组成, 一个用于样品, 另一个用于参考, 该支架安装在分光光度计中, 在仪器的前面板下方便地将实用程序路由。使用菜单按钮操作 tc 半温度控制器, 或通过可选程序 t-app 进行控制。

Product Features 产品特征

Temperature control for both sample and reference 样品和参考的温度控制
Variable speed magnetic stirring变速磁搅拌
Dry gas purge for work at low temperature  低温工作用干气净化
Bases and posts for mounting in the spectrophotometer 分光光度计中的安装基座和立柱
Cuvette lifters for removal of cuvettes  用于去除切割机的刮板提升机
Probe input for sample temperature 样品温度的探头输入
NIST-traceable temperature calibration nist 可追溯温度校准

产品参数:

Number of cuvettes controlled: 2

Typical use: Absorbance (sample and reference)

Standard cuvette size (outside) :12.5 mm x 12.5 mm

Optical ports per cuvette: 2

Optical port dimensions:12 mm high x 10 mm wide

Windowed jacket available: No

Factory set temperature range: -40 °C to +110 °C

Temperature range easily achievable: -15 °C to +110 °C

Temperature precision: ±0.02 °C

Temperature accuracy: Better than ±0.15 °C from 0 ⁰C to +80 ⁰C

Accuracy full range: Better than ±0.3 ⁰C from ‐15 ⁰C to +110 ⁰C

Temperature reproducibility: Better than ±0.07 °C

Thermistor probes accepted: Series 400 or Series 500

Magnetic stirring motor type: DC

Magnetic stirring speed range: 100-2500 rpm

Default stirring speed: 1200 rpm

Stirring speed best performance: 200-1800 rpm

 

Basic:

Packaget2x2 Sport – Pair of cuvette holders

TC 1/Dual – Temperature Controller

Base plates, posts (standard z-height = 15 mm)

BATH 10 – Circulator

NIST-traceable calibration and performance data

Accessories – Two magnetic stir bars, tubing and cables

 

Accessories:

T-App

Water Circulators

Cuvette Adapters

Thermistor Probes

 