Luma 40/EVO 支架 光学支架 精密温控样品池支架

温度控制的立方座, 具有四个光学端口, 用于 Evolution 300和600分光光度计

The Luma 40/EVO has four optical ports around the cuvette. Use it for absorbance measurements when your experiment requires access at 90 degrees for photochemical activation. Purchase the extended temperature option and run the Luma 40/EVO up to 150 °C for non-aqueous samples. Add a Windowed Jacket, purge with dry gas and run the temperature down to at least -40 °C. The Luma 40 is also ideal for work in the biological range of 0 °C to 110 °C. Use it to observe absorbance spectra at precisely maintained temperatures, or for temperature ramping, to observe how absorbance changes with temperature. Operate the TC 1/Single Temperature Controller using the menu button or control via optional program T-App.
luma 40/evo 在 cuvette 周围有四个光学端口。当您的实验需要在90度的温度下进行光化学激活时, 使用它进行吸收测量。购买扩展温度选项, 并运行 luma 40/evo 高达150°c 的非水样。加入一个风夹克, 用干气清洗, 并将温度降至-40°c。luma 40 也是在0°c 至110°c 的生物范围内工作的理想选择。使用它来观察在精确保持温度下的吸收光谱, 或在温度下倾斜, 观察吸收率如何随温度变化。使用菜单按钮操作 tc 半温度控制器, 或通过可选程序 t-app 进行控制。

Product Features产品特点

Four optical ports 四个光学端口
Rapid, precise temperature control 快速、精确的温度控制
Optional extended temperature range 可选的扩展温度范围
Windowed jacket available for temperature extremes 可用于极端温度的风装夹克
Variable speed magnetic stirring 变速磁搅拌
Dry gas purge for work at low temperature 低温工作用干气净化
Base and post for mounting in spectrometer 光谱仪安装的基础和立柱
Probe input for sample temperature 样品温度的探头输入
NIST-traceable temperature calibration nist 可追溯温度校准
Default stirring speed: 1200 rpm 默认搅拌速度: 1200 转/分
Stirring speed best performance: 900-1800 rpm 搅拌速度最佳性能: 900-1800 rpm

Technical Specifications

Number of cuvettes controlled:1
Typical use:Fluorescence
Standard cuvette size (outside):12.5 mm x 12.5 mm
Optical ports per cuvette:4
Optical port dimensions:12 mm high x 10 mm wide
Built-in optical slit holders:4 per cuvette
Cuvette z-height:8.5 or 15 mm for most models
Windowed jacket available:2, 3 or 4 windows
Normal factory set temperature range:-40 °C to +105 °C
Temperature range easily achievable:-15 °C to +105 °C
Range with special techniques:-40 °C to +105 °C
Extended factory set temperature range:-55 °C to +150 °C
Extended temperature range achievable:-50 °C to +150 °C
Temperature precision:±0.02 °C
Temperature accuracy:Better than ±0.15 °C from 0 ⁰C to +80 ⁰C
Accuracy greater range:Better than ±0.3 ⁰C from ‐20 ⁰C to +110 ⁰C
Temperature reproducibility:Better than ±0.07 °C
Thermistor probes accepted:Series 400 or Series 500
Magnetic stirring motor type:Stepper
Magnetic stirring speed range:1-2500 rpm
Default stirring speed:1200 rpm
Stirring speed best performance:60-1800 rpm

Basic Package
Luma 40 – Temperature-controlled cuvette holder (standard z-height = 15 mm)
TC 1/Single – Temperature Controller
Base plate
Bath 10 – Circulator
NIST-traceable calibration and performance data
Accessories – Magnetic stir bar, tubing, cables, optical slits and opaque cap