HumiSys XR相对湿度发生器

The Humisys XR RH generator featuring fully automatic and manual mode was designed for extending full range of RH generation down to 0 C and allowing larger range of RH in subzero C region then it is commonly available using other humidity generatores. Combination of of two-temperature and divided-flow techniques is emploeyd. Embedding of precision RH probe in the condensor housing that controlls the temperature of the humidified stream from the first stage of the two-temperature method allows for generating very low dew points in combination of the divided flow technique in the second stage. External RH probe and external temperature probe (RTD based) are very useful in easy interfacing with other analytical instruments.

Design Features:
-Combination of two-temperature and divided flow techniques
(two mass flow controllers)
-Extended RH spectrum in the near zero and sub-zero C
temperatures region
-Upper temperature  60C for standard models
-Flow rate –  typically up to 1 L/min
-Heated transfer hose
-External temperature probe (RTD based)
-Automatic water supply system
-Auto/Manual operation
-Automatic water supply system
-Precision pressure regulator and gauge
-Open design for easy interfacing with other equipment
-Internal and external RH probe (RH and temperature data)
-Auto/Manual operation
-USB or serial port for data transfer
-Windows based software for experiment design and control
(Our software is compatible with 64-bit Windows® system)
-Very low cost


Related posts