Beam Imaging 电子束观察系统 BOS-75

In standard form, the model BOS-75 Beam Observation system has a 75mm diameter active area MCP and P-43 phosphor screen assembly, attached to a 8.00″ (203mm) conflat flange with glass viewport. Electrical connections to the phosphor screen and MCP are made with MHV feedthroughs.

Optional configurations are discussed in detail below and include, dual MCP plates (chevron), SHV feedthroughs, and CCD camera systems.

Imaging Area:

  • 75mm Diameter

MCP: (Standard)

  • 3.410″ Diam. (BOS-75), 25 micron channel diameter, Detection or Imaging Grade
  • 75mm active area, 32 micron pitch, 8° Bias Angle, 40:1 Aspect Ratio (Standard, BOS-75)
  • Min. Gain:  1 x 10^4 (single plate, Std.) 1000V  > 107 (chevron, OPT-01), 2000V

Phosphor Screen (Standard)

  • P-43 with aluminum overcoat. P-43 Peak Wavelength: λ= 545 nm.

Power Supply Requirements:

  • 0 – + 1000V, 1mA single MCP (Standard)
  • 0 – + 2000V, 1mA dual MCP (OPT-01)
  • 0 – + 5000V, 1 mA Phosphor Screen

Beam Energy Range:

  • 1 eV to over 50 keV

Beam Current Range:

  • < 10 μA (with optional beam attenuation grids)

Vacuum:

  • 1 x 10-6 Torr or better required to operate MCP
  • UHV compatible, maximum bakeout temperature 300 C

Welded Glass Windows (Standard)

  • A non-removable glass window that is welded to the flange is standard. This allows bakeout temperatures > 200C.