SMART CHARGE 5 TRACER GAS CHARGING SYSTEM 追踪气体充气系统
The Smart Charge 5 is a fully automatic, custom configurable evacuation and tracer gas backfill system with built-in gross leak test. The system is ideally suited for pre-charging of product with tracer gas, such as Helium, Hydrogen or others prior to the leak test processes. It can also be used with any gas charging application that requires evacuation and subsequent pressurization of a test product.
Smart Charge 5是一款全自动、可自定义配置的排空和示踪气体充气系统,内置总泄漏测试。该系统非常适合在泄漏测试过程之前用示踪气体(如氦气、氢气或其他气体)预充产品。它也可以用于任何需要排空和随后对测试产品加压的充气应用。
Main Features
- Fastest response time in the industry (less than 0.5 seconds)
- Tracer gas background suppression
- Sensitivity to 5 x 10-8 atm cc/secs
- Less than 3 minutes total start-up time
- Fully automatic calibration sequence with built-in dual purpose gas leak
- User selectable between helium and hydrogen sniffer
- Front panel key lock-out
- Sealed membrane keypad
- Fully interactive system diagnostics with multilingual displays
- High output (110dB) adjustable audio alarm业界最快的响应时间(小于0.5秒)
示踪气体背景抑制
灵敏度为5 x 10-8 atm cc/sec
总启动时间小于3分钟
内置双重用途气体泄漏的全自动校准序列
用户可选择氦气和氢气嗅探器
前面板钥匙锁定
密封膜键盘
具有多语言显示屏的完全交互式系统诊断
高输出(110dB)可调音频报警器
Flexible Communication Format
- Option Full duplex RS232 communication
- Option AC & DC PLC interfacing
- Option Analog outputs
- Option Data Link Software选配全双工RS232通信
选配交流和直流PLC接口
选配模拟输出
选配数据链路软件
SMART CHARGE 5: CHARACTERISTICS
Options
- Analog
- User I/O
- Printer
- Low pressure feature, 0-20 psig
- Nitrogen charge up 700 psig
- Barcode scanner feature

TECHNICAL SPECIFICATIONS |
---|
Power | Universal Voltage 110 vac or 220-Volt AC, 50 – 60 Hz, 6.5 amps nominal |
---|---|
Dimensions (WxHxD) | 344 x 395 x 618 mm 13.5″ x 15.5″ x 24.4″ inches |
Utility Fittings | 1⁄4” NPT female 1⁄2” NPT female for Part Evacuation |
Pressure Transducer | Resolution of 0.01 psi in vacuum mode, or in pressure mode. Accuracy of ± 0.25%. |
Valving | Fast acting pull type solenoid featuring epoxy encapsulated design |
Manifolding | Seamless tubing with compression fitting interconnects. Helium leak tested to 1.0 × 10-6 atm-cc/sec @ 700 psig |
Product Evacuation Pump | Air Driven Venturi Pump Providing ultimate vacuum level 25.5″ Hg |
Electronics | Microprocessor-based programmable logic controller |
Gauging | Digital part pressure and leak rate displayed on operator interface |