MR 1012 S 数字欧姆表

Owing to its high measuring rate, its compact and robust design, ist accuracy and long-term stability, the digital ohmmeter MR1012S can be used in applications with high requirements concerning the measuring stability and reliability. This is guaranteed by the integrating measuring method according to the quotient principle, the compensation of all thermoelectric forces and other offset voltages, as well as the suppression of interference voltages in the measuring and data lines induced e.g. by HF.

数字欧姆表MR1012S具有测量速率高、结构紧凑、设计可靠、精度高、长期稳定等优点,可用于对测量稳定性和可靠性提出较高要求的应用。根据商原理的积分测量方法、所有热电力和其他偏置电压的补偿以及抑制测量和数据线中的干扰电压,例如高频引起的干扰电压,都能保证这一点。

Resistance measurement
Range 10 mΩ – 100.00 kΩ, decadic
Overrange +80 %, to 18000
Resolution 1 ;@Range 1 mΩ
Condition for Error Specs @ 10 Measurements / s ; 20/23°C
Max. Measurement Error ± 0.03 % RDG ± 0.02 % RNG
Additonal Temperature Error ± 0.0002 % x (|t| – 20/23°C) RDG
Long term Stability ± 0.002 % / Year RDG
Measuring Method integrating dual slope quotient
Object Connection 4-Pole due to Kelvin
Current 100 ;A (10 kΩ) to 1 A (1, 10 mΩ)
Max. Lead Resistance 2Ω (tested), 3Ω max. @ 1A
Range selection automatically, using keypad, via RS232
Display LED, 3 ½ or 4 ½ digits, selectable
Speed 5, 10, 20 Measurements per second available
Temperature measurement
Programmable Cycles every 1st to 1000thmeasurement
Reference conversion to 20 °C or 23 °C
Temeprature Coefficient – 9.99 to + 9.99 ‰/K via RS232 or Keypad
Inputs PT 100 probe, or via keypad or RS232
Error detection prior to EVERY single measurement
Current Connection Errors display: ‚CUR‘, RS232: ‚ECUR‘
Sense Connection Errors display: ‚SEN‘, RS232: ‚ESEN‘
Overrange > 80% display: ‚OVL‘, RS232: ‚EOVL‘
Thermoelectric Force Compes. prior to every single cycle, automatically
Limit Values
Input using Keypad, via RS232
Off-Limit Condition visible by LEDs, via RS232, via PLC
Start of Measurment using Keypad, via RS232 (pot.free)
and IEEE – 488, via PLC (potential free contact),
via Foot Switch (optional)
Interfaces RS232C (full device control)
PLC (<,=,>,EOC,GO,REM)
Printer (parallel, ANSI standard)
IEEE – 488 (optional)
Environment Operating: 0° – +50°C, RHD 80% max.
Storage: -20°C … +70°C, both not condensing
EMC IEC801-2/-3 , EN55011
Dimensions 260 x 80 x 240 mm (WxHxD)
Weight approx. 3 kg
Multiplexer within the normal dimensions:
This instrument can be expanded to 6/10
measuring channels, 10 with common ground
(electronic switches), 6 all separate with six 4-
pole relais (Options).
For more channels (up to 50) see our MR1012MP.