Lambda-messtechnik防护热板法导热系数仪EP500e、热导率测试仪EP500e
技术规格
Parameter | Version A | Version C |
Real Measurement Range is the Thermal Transmittance Resistance of the inserted sample (correlated with the measurement of the electrical heating power of the measuring heater): | R =0.1 … 14 m²K/W | R = 0.025 … 14 m²K/W |
corresponding Thermal Conductivity (ca.): | λ = 0 … 0.5 W/(m·K) | λ = 0 … 3 W/(m·K) |
Measuring Area/Zone: | 200 x 200 mm² | 150 x 150 mm² |
Sample Thickness: | 10 … 120 mm (optional 4 … 200 mm) | |
Sample Dimensions: |
| |
Reproducibility: | < 0.5 % (mostly < 0,2 %) | |
Accuracy: | < 1.0 % (mostly < 0,7 %) | |
Measurement Temperature Range (Standard): | 10 … 40 °C (optional -10 … 50 °C) | |
Temperature Difference Range (Standard): | 5 … 15 K (optional 20 K) |