Technical Specifications Incubator

Incubator type Single Plate Incubator MP Single Plate Incubator DWP
Input voltage 24 Vdc
Input power Max 30 W Max 45 W
Maximum numbers of devices per pow-er supply (2400121) 6 4
Temperature range Room temperature +5 K up to +80 °C (+175 °F)
Tolerable relative humidity ambient 10-80% relative at +20 °C (+68 °F) up to +30 °C (+86 °F)
Maximum load of the shaker platform N.A.
Shaker frequency range N.A.
Maximum shaking amplitude (difference between the maxima) N.A.
Shaking direction N.A.
Dimension in YModule w connector cable 295.5 mm
Dimension in YModule wo connector cable 268.5 mm
Dimension in XModule 149 mm
Dimension in ZModule wo pin 58 mm 104 mm
Open drawer incl. door YDrawer 193.5 mm 223.5 mm
Dimension of the thermal chamber in ZChamber 18 mm 50 mm
Weight 2.5 kg 3.4 kg
Maximum number of devices in one stack due to interaction during shaking and power limitation of power supply 5 4
Drawer specifications
Accuracy of the end position of the drawer in x, y and z (drawer open); single unit, all parts at specified ambient temperature ±0,30 mm
Maximum drawer load 0,25 kg
Open drawer resistance for a force in Z-axis without plastic deformation 10 N
Environmental conditions
Tolerable relative humidity:
Operation 10-80% relative at+20 °C (+68 °F) up to +30 °C (+86 °F)
Transportation and storage -10 °C to +60 °C (+14 °F to +140 °F)
Temperature:
Operation +20 °C to + 30 °C (+68 °F to +86 °F)
Transportation and storage -10 °C to + 60 °C (+14 °F) to 140 °F)

Shaker functionality Incubator Shaker MP / DWP

Parameter Value
Amplitude range 0.0 – 3.0 mm *
Speed range 6.6 – 30.0 Hz *
Speed precision ±2 Hz
Load 40 – 250 g * (weight of plate and liquid and adapter)
Supervision Speed and amplitude (closed loop controller)
Shaker movement North – South (linear)
East – West (linear)
Round (orbital)
“Eight” (this is possible but not recommended because it might damage the shaker)
Shaker movement
Amplitude accuracy** ±20% (of the target amplitude)
Phase shift accuracy** ±11 °
** Validated for the Heating Plate movement / average value over 10 cycles
Single Axis (North – South and East – West) Shaker Load: 100 g
Shaker Shaker amplitude [mm]
Speed [Hz] Single device Maximum stack
10 X = 1.0 to 2.2 or Y = 1.0 to 2.2 X = 1,0 to 1.9 or Y = 1.0 to 1.9
15 X = 1.0 to 2.7 or Y = 1.0 to 2.7 X = 1.0 to 2.4 or Y = 1,0 to 2.4
20 X = 1,0 to 1.2 or Y = 1.0 to 1.2 X = 1.0 to 1.2 or Y= 1.0 to 1.2
Round (Phase shift = 90 °) Shaker load: 100 g
Shaker Shaker amplitude [mm]
Speed [Hz] Single device Maximum stack
10 X = 1.0 to 2.2
Y = 1.0 to 2.2
X = 1,0 to 1.9
Y = 1.0 to 1.9
15 X = 1.0 to 2.7
Y = 1.0 to 2.7
X = 1.0 to 2.4
Y = 1.0 to 2.4
20 X = 1.0 to 1.2
Y = 1.0 to 1.2
X = 1.0 to 1.2
Y = 1.0 to 1.2
„Eight“ (Phase shift = 0°) Shaker load: 100 g
Shaker Shaker amplitude [mm]
Speed [Hz] Single device Triple stack
X = 12 / Y = 24 X = 1.0 to 2.4 / Y = 1.0 to 2.4 X = 1.0 to 2.2 / Y = 1.0 to 2.2
X = 15 / Y = 30 X = 1.0 / Y = 1.0 X = 1.0 / Y = 1.0

Thermal functionality for all Incubators

Parameter Value
Temperature range Room temperature plus 5 K to +80 °C [+176 °F]
Accuracy absolute (@ ambient temperature stability ±1 K) The accuracy is defined as the difference between the target temperature of the heating plate and the average of the measured temperature on the heating plate. The heating plate temperature is measured at 5 positions on the heating plate according to the Inheco measurement head PM-0105. The 5 measurement positions are according to the Plate positions A1, H1, E6, A12 and H12.

Target temp. validation tool Accuracy [°C] for all incubator
+37 °C [+98,6 °F] ±1.0
+60 °C [+140 °F] ±1.5
+80 °C [+176 °F] ±2.0
Uniformity measured in well Uniformity is defined by the maximum difference between lowest and highest temperature according to a Greiner 96 Flat Bottom Plate (postitions A1, A12, E7, H12)

Target temp. validation tool Uniformity [°C] for non shaking devices Uniformity [°C] for shaking devices
+37 °C [+98,6 °F] ≤ 0.8 ≤ 0.8
Evaporation in unsealed Micro Plates (can be reduced by using lid or cover)
Incubator non shaker temperature +37 °C [+98,6 °F], max. 25% loss of liquid in 4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level Incubator shaker temperature +37 °C [+98,6 °F], max. 32% loss of liquid in 4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level
Heat-up time at start temperature +20°C [+68°F] with target temperature +37°C [+98°F]
Max. 30 min; Volume 280 µl per well. * Max. 100 min; Volume 1.5 ml per well. ** Max. 30 min; Volume 280 µl per well. * Max. 120 min; Volume 1.5 ml per well. **

* The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1) Greiner 96 Flat Bottom Plate

** The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1) NUNC 96 Deep Well Plate