Inheco单板培养箱、孵育器、单板培养箱振荡器、Inheco孵化机摇床、培养箱、Inheco孵化器
为细胞培养提供出色的加热和保护
增加自动化细胞培养培养的产量,而不占用宝贵的台面空间。Inheco单板培养箱小巧灵活,可快速达到工作温度。
恒温箱提供受保护的加热环境,细胞样本可以在最佳条件下生长,避免光照和污染物。
节省空间
Inheco孵化器位于甲板外或甲板旁边,以节省您的空间。它通过自己的内部控制器进行调节,实现甲板上的自动化应用。其自动打开和关闭的门将样品板移出,即使只有该部分伸入液体处理平台的工作区域,也能实现甲板上的自动化应用。
快速准确
Inheco单板培养箱只需10分钟即可达到工作温度。为了向您提供高度精确的分析,我们的培养箱最高工作温度为80°C,温度精度为0.3 K,均匀性为0.5 K。
多才多艺的
微孔板有两种型号可供选择:仅从底部加热,适用于有盖微孔板;深孔:从各个方向加热,以避免冷凝,适用于样品瓶和容器。为了获得更大的灵活性,可以将两个或更多版本堆叠在一起。所有这些都在149毫米x 268毫米的工作站空间内。放置在最适合您的设置的位置–您可以将保育箱放置在不同的方向,以便于插入和取出孔板,并且无需从抽屉中取出孔板即可完成移液。
易于集成
API可用,并且已经与主要液体处理系统集成。对于堆叠设备,设备是单独和独立控制的(主从原则)。
Technical Specifications Incubator
Incubator type |
Single Plate Incubator MP |
Single Plate Incubator DWP |
Input voltage |
24 Vdc |
Input power |
Max 30 W |
Max 45 W |
Maximum numbers of devices per pow-er supply (2400121) |
6 |
4 |
Temperature range |
Room temperature +5 K up to +80 °C (+175 °F) |
Tolerable relative humidity ambient |
10-80% relative at +20 °C (+68 °F) up to +30 °C (+86 °F) |
Maximum load of the shaker platform |
N.A. |
Shaker frequency range |
N.A. |
Maximum shaking amplitude (difference between the maxima) |
N.A. |
Shaking direction |
N.A. |
Dimension in YModule w connector cable |
295.5 mm |
Dimension in YModule wo connector cable |
268.5 mm |
Dimension in XModule |
149 mm |
Dimension in ZModule wo pin |
58 mm |
104 mm |
Open drawer incl. door YDrawer |
193.5 mm |
223.5 mm |
Dimension of the thermal chamber in ZChamber |
18 mm |
50 mm |
Weight |
2.5 kg |
3.4 kg |
Maximum number of devices in one stack due to interaction during shaking and power limitation of power supply |
5 |
4 |
Drawer specifications |
Accuracy of the end position of the drawer in x, y and z (drawer open); single unit, all parts at specified ambient temperature |
±0,30 mm |
Maximum drawer load |
0,25 kg |
Open drawer resistance for a force in Z-axis without plastic deformation |
10 N |
Environmental conditions |
Tolerable relative humidity: |
Operation |
10-80% relative at+20 °C (+68 °F) up to +30 °C (+86 °F) |
Transportation and storage |
-10 °C to +60 °C (+14 °F to +140 °F) |
Temperature: |
Operation |
+20 °C to + 30 °C (+68 °F to +86 °F) |
Transportation and storage |
-10 °C to + 60 °C (+14 °F) to 140 °F) |
Shaker functionality Incubator Shaker MP / DWP
Parameter |
Value |
Amplitude range |
0.0 – 3.0 mm * |
Speed range |
6.6 – 30.0 Hz * |
Speed precision |
±2 Hz |
Load |
40 – 250 g * (weight of plate and liquid and adapter) |
Supervision |
Speed and amplitude (closed loop controller) |
Shaker movement |
North – South (linear)
East – West (linear)
Round (orbital)
“Eight” (this is possible but not recommended because it might damage the shaker) |
Shaker movement |
Amplitude accuracy** ±20% (of the target amplitude)
Phase shift accuracy** ±11 °
** Validated for the Heating Plate movement / average value over 10 cycles |
Single Axis (North – South and East – West) Shaker Load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Maximum stack |
10 |
X = 1.0 to 2.2 or Y = 1.0 to 2.2 |
X = 1,0 to 1.9 or Y = 1.0 to 1.9 |
15 |
X = 1.0 to 2.7 or Y = 1.0 to 2.7 |
X = 1.0 to 2.4 or Y = 1,0 to 2.4 |
20 |
X = 1,0 to 1.2 or Y = 1.0 to 1.2 |
X = 1.0 to 1.2 or Y= 1.0 to 1.2 |
Round (Phase shift = 90 °) |
Shaker load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Maximum stack |
10 |
X = 1.0 to 2.2
Y = 1.0 to 2.2 |
X = 1,0 to 1.9
Y = 1.0 to 1.9 |
15 |
X = 1.0 to 2.7
Y = 1.0 to 2.7 |
X = 1.0 to 2.4
Y = 1.0 to 2.4 |
20 |
X = 1.0 to 1.2
Y = 1.0 to 1.2 |
X = 1.0 to 1.2
Y = 1.0 to 1.2 |
„Eight“ (Phase shift = 0°) |
Shaker load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Triple stack |
X = 12 / Y = 24 |
X = 1.0 to 2.4 / Y = 1.0 to 2.4 |
X = 1.0 to 2.2 / Y = 1.0 to 2.2 |
X = 15 / Y = 30 |
X = 1.0 / Y = 1.0 |
X = 1.0 / Y = 1.0 |
|
Thermal functionality for all Incubators
Parameter |
Value |
Temperature range |
Room temperature plus 5 K to +80 °C [+176 °F] |
Accuracy absolute (@ ambient temperature stability ±1 K) |
The accuracy is defined as the difference between the target temperature of the heating plate and the average of the measured temperature on the heating plate. The heating plate temperature is measured at 5 positions on the heating plate according to the Inheco measurement head PM-0105. The 5 measurement positions are according to the Plate positions A1, H1, E6, A12 and H12.
Target temp. validation tool |
Accuracy [°C] for all incubator |
+37 °C [+98,6 °F] |
±1.0 |
+60 °C [+140 °F] |
±1.5 |
+80 °C [+176 °F] |
±2.0 |
|
Uniformity measured in well |
Uniformity is defined by the maximum difference between lowest and highest temperature according to a Greiner 96 Flat Bottom Plate (postitions A1, A12, E7, H12)
Target temp. validation tool |
Uniformity [°C] for non shaking devices |
Uniformity [°C] for shaking devices |
+37 °C [+98,6 °F] |
≤ 0.8 |
≤ 0.8 |
|
Evaporation in unsealed Micro Plates (can be reduced by using lid or cover) |
Incubator non shaker temperature +37 °C [+98,6 °F], max. 25% loss of liquid in 4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level |
Incubator shaker temperature +37 °C [+98,6 °F], max. 32% loss of liquid in 4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level |
|
Heat-up time at start temperature +20°C [+68°F] with target temperature +37°C [+98°F] |
Max. 30 min; Volume 280 µl per well. * |
Max. 100 min; Volume 1.5 ml per well. ** |
Max. 30 min; Volume 280 µl per well. * |
Max. 120 min; Volume 1.5 ml per well. ** |
|
* The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1) Greiner 96 Flat Bottom Plate
** The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1) NUNC 96 Deep Well Plate