Thyracont Vacuum Instruments GmbH Vacuum Transducer 真空传感器 VSP63MV
DN16KF 0-10V output
MEASURING PRINCIPLE Heat conduction (Impulse Pirani), depending on gas type
MATERIALS IN CONTACT WITH VACUUM Stainless steel 1.4307, nickel, tungsten, glass
MEASUREMENT RANGE 1000 – 1e-4 mbar (750 – 1e-4 Torr), max. overpressure 10 bar absolute
REPEATABILITY 2 %
REACTION TIME Max. 200 ms
POWER SUPPLY 15 – 30 VDC
ELECTRICAL CONNECTION Hirschmann, 6 pole, male, lockable
POWER CONSUMPTION Approx. 10 mA
OPERATING TEMPERATURE 5 °C…+60°C
STORAGE TEMPERATURE -20…+70 °C
MAX. BAKE OUT TEMPERATURE 150°C at the flange
VACUUM CONNECTION DN 16 ISO-KF
PROTECTION CLASS IP40
Thyracont Vacuum Instruments GmbH Vacuum Transducer 真空传感器 VSP63MV
询价采购Thyracont Vacuum Instruments GmbH Vacuum Transduce
请微信扫描下方二维码或手动添加微信号2351992198>