TAURUS instruments GmbH Measuring and Control - Burning Control Measuring and Control - Burning Control测量单元燃烧控制 SBI.CONTROL
Technical Specification
All modules for control and data acquisition for the SBI test are indicated by a central controller
controlled and evaluated by means of the operation of the priority SBI control station CS occurs.
It is also possible directly on the touch screen operations to control, display measured values
and perform service functions. The controller firmware can be updated, so future demands on
the system can be implemented without changes to the module components.
Controller SBI20_CPU
Windows CE Controller
Colour Touch Screen 7“, 840 x 480 pixels
ARM9 Processor 400 MHz
64 MB Flash, 64 MB RAM
2 x USB, 1x Ethernet
Windows CE 2012 7.0
Temperature
Measuring SBI20_AIN
4 x Thermal voltage inputs
24 bit Sigma Delta integrating ADC
Measuring range -100 – 1300 °C, resolution 0.1K
Comparison point compensation
Control of
MFC SBI20_MFC
1 x digital output (RS232) for MFC1
1 x digital output (RS232) for MFC2
Control of
Gas valves SBI20_FET
8 x switched outputs to the following components
Propane valves, main burner, auxiliary burner
Calibration gas and zero gas valves
Measuring gas valve and measuring gas pump
Peltier cooler/
Flame control SBI20_PEK
2 x digital input, flame control
1 x Control Peltier cooler, gas conditioning
Control of
Gas analysis SBI20_UOX
2 x RS485, control / service
Converter Harth- to Modbus
Serial
Interfaces SBI20_SER
1 x RS232, SBI control station
1 x RS485, control panel
Differential
Pressure/Light SBI20_DRL
1 x analog input differential pressure
1 x analog input differential pressure 2 (optional)
1 x digital input (RS485) for measuring light receiver
1 x digital input (RS485) for ambient humidity / pressure
1 x constant current source for measuring light transmitter
1 x voltage output to control the deduction system
16 bit ADC, measuring range +/- 10V, resolution 1mV
High constant current source for lamp 5V/2A
Light Receiver SBI20_LME
Silicon photo detector
Tempered, heat-proof optics
Spectral filter allocation of CIE
Measuring light amplifier, AD converter 16 bit
DIN socket, serial interface RS485
Dimensions (L x Ø) 190 x 40 mm
Aluminum housing with mounting flange, black anodized
Light Transmitter SBI20_LMG
Halogen spot light source 10W, 2900K color temperature
Tempered, heat-proof optics
Beam diameter 25mm, d / f = 0.0375
DIN socket for power supply (constant current from module)
Dimensions (L x Ø) 190 x 40 mm
Aluminum housing with mounting flange, black anodized
Gas Processing SBI20_MGA
Roughly particulate filter mounted on the measuring tube
Finely particulate filter with glass fiber filter element 0.1
Gas pump 60-200 l / min
Measurement gas Peltier cooler to -10 ° C with condensate drain
and collecting vessel
Flow indicator inlet 0-200 l / min
Flow indicator outlet 0-100 l / min
Bypass valves to adjust the measuring gas flow through the analyzer
Solenoid valves for control of zero gas, calibration gas, measurement gas
Reservoir for dry gas, plug-in
Analyzer SIEMENS ULTRAMAT/OXYMAT 6E
Measuring components O2, CO2
Display in vol%
Paramagnetic O2 measurement according to
standard
Infrared CO2 measurement according to
standard
1 x synthetic air reference gas
Performance monitoring and reporting
19 "rack, 3U
Power supply 230 V, 50 Hz
Overall Design 19“- Industrial Cabinet
Wired and hosed systems
Colour gray / light gray
Protection class IP20
Dimension (W x D x H) 65 x 65 x 175 cm
Weight 168 kg
Included Measuring Unit SBI Burning Control
Connection cable
English Manual
Options Optical filter set
Zero gas N2, calibration gas 20% O2 / 5% CO2 / N2, 20l bottle
Synthetic air reference gas, 20l bottle
Hand anemometer to determine the flow profile
Holder / assistance for measuring anemometer
Heptane to heptanes test
Filter, fine filter replacement
Scale for test mass flow controller
TAURUS instruments GmbH Measuring and Control - Burning Control Measuring and Control - Burning Control测量单元燃烧控制 SBI.CONTROL
询价采购TAURUS instruments GmbH Measuring and Control - Bu