The Bionetics Corporation, formerly Intek, Inc. 210R-XDT-TU1/4(C20)(SS)(350F)(1"#150RF)-JB(SS)-HART-24V-ISB-5M 液体流量计 210R-XDT-TU1/4(C20)(SS)(350F)(1-150RF)
Consisting of flow sensing transducer, separate electronics package, and interconnect cable
Flow Data
Fluid: 75% Phosphoric Acid (see Note 1)
Customer Line Size: 1/4-inch
Customer Flow Range: 0 - 15 kg/hr
Calibrated Flow Range: 1.5 - 15 kg/hr
Temperature - Max Requirement: 163℃
Temperature - Operating Range: 70 - 70℃
Pressure - Max Requirement: 1.03 MPa-g
Pressure - Operating Range: 0.79 MPa-g
Fluid Viscosity: 21.5 cP
Fluid Density: 1519 kg/m^3
Instrument Data
Transducer Type: Straight tube
Wetted Material: Alloy 20 (Carpenter 20) (see Note 2)
Shell Material: SS 316
Process Connection: ASME B16.5 Raised Face Flange #150: 1" Alloy 20
Local Display: 2x16 character LCD with selectable Flow Rate, Totalizer and Fluid Temperature
Calibration Units: kg/hr.
Outputs: Passive 4 to 20 mA with HART
Input Power: 24VDC
Electronics Enclosure: NEMA 7 10x14x6 inches. Electronics are in a flame proof enclosure, approved for IECEX EXd IIB IP66 areas.
Cable Length: 5M; (付费可升级至200 ft)
Sensor tube size: 1/4-inch outside diameter tube with 0.049-inch tube wall, and 0.152 inch inside diameter. 13 inches end to end.
Min. | Nom. | Max. | |
Measuring Range: | 1.5 kg/hr | 7.5 kg/hr | 15 kg/hr |
Velocity: | 0.0222 m/s | 0.111 m/s | 0.222 m/s |
Measurement Uncertainty: | 1.04% | 1.29% | 1.44% |
Pressure Drop: | 0.05 psi | 0.25 psi | 0.5 psi |
Notes
1. Fluid - Calibration will be performed on the actual fluid if it can be handled safely and is readily available or if it can be supplied by Buyer. Otherwise, a substitute calibration fluid will be used, in which case a field calibration adjustment may be necessary to guarantee the highest levels of accuracy.
2. Wetted Material - The user is responsible for the selection of process-wetted materials in view of their corrosion resistance. Intek, Inc. makes no guarantees and assumes no liability for the corrosion resistance of the materials selected here for the application described above.
3. ISB - The electronics and sensor listed in item 1 above are approved hazardous environments, sensor meets ATEX certification for Ex ia IIC T4 Ga areas, enclosure meets ATEX approved for IECEX EXd IIB IP66 areas.
4. Transducer design temperature - 163C Max., ATEX approval valid to temperatures less than or equal to 120C